SensitronProducts / Sensitron / SMART 3G D2

Sensitron / Gas detection
SMART 3G D2
Display-equipped industrial gas detector family for ATEX Zone 1. Magnetic non-intrusive calibration, local gas indication and sensor-specific variants support commissioning and maintenance.
| Manufacturer | Sensitron |
|---|---|
| Model / family | SMART 3G D2 |
| System | Gas detection |
| Selection & quotation | Based on project specification, configuration and quantities. |
Technical specifications
| Operating voltage | 12-24 VdcSource · 2 ↗ |
|---|---|
| IP rating | IP65Source · 3 ↗ |
| Application — Catalytic | Industrial area, ATEX Zone 1 and Zone 21 (when equipped with dust adapter, optional)Source · 2 ↗ |
| Application — Electrochemical | Industrial area, ATEX Zone 1 and Zone 21 (when equipped with dust adapter, optional)Source · 2 ↗ |
| Application — Intrinsically safe EC | Industrial area, ATEX Zone 1 and Zone 21 (when equipped with dust adapter, optional)Source · 2 ↗ |
| Application — Infrared | Industrial area, ATEX Zone 1 and Zone 21 (when equipped with dust adapter, optional)Source · 2 ↗ |
| Application — PID | Industrial area, ATEX Zone 1 e Zone 21 (when equipped with dust adapter, optional)Source · 2 ↗ |
| Detection principle — Catalytic | CatalyticSource · 2 ↗ |
| Detection principle — Electrochemical | Electrochemical cellSource · 2 ↗ |
| Detection principle — Intrinsically safe EC | Electrochemical cellSource · 2 ↗ |
| Detection principle — Infrared | InfraredSource · 2 ↗ |
| Detection principle — PID | Photoionizator (PID)Source · 2 ↗ |
| Detectable gases — Catalytic | Flammable gasSource · 2 ↗ |
| Detectable gases — Electrochemical | O2 [0-25% Vol], CO [0-500/1000 ppm], NH3 [0-100/200/1000/5000 ppm], H2S [0- 50/100 ppm], NO [0 - 100 ppm]; H2 [0 - 2000/3000 ppm]Source · 2 ↗ |
| Detectable gases — Intrinsically safe EC | Sulphur dioxide [0-20 ppm] Nitrogen dioxide [0-20 ppm] Chlorine [0-5 ppm]; [0-20 ppm]Source · 2 ↗ |
| Detectable gases — Infrared | Flammable, refrigerant and CO2Source · 2 ↗ |
| Detectable gases — PID | Volatile Organic Compounds (VOC) having photoionization potential up to 10.6 eVSource · 2 ↗ |
| Measurement range — Catalytic | 0 - 100% LFLSource · 2 ↗ |
| Measurement range — PID | 0 - 20 ppm (Isobutylene) 0 - 200 ppm (Isobutylene) 0 - 2000 ppm (Isobutylene)Source · 2 ↗ |
| Outputs — Catalytic | Analogue 4-20 mA output; 3 relay board; serial RS485 output (optional)Source · 2 ↗ |
| Outputs — Electrochemical | Analogue 4-20 mA output; 3 relay board; serial RS485 output (optional)Source · 2 ↗ |
| Outputs — Intrinsically safe EC | Analogue 4-20 mA output; 1 or 3 relay board (optional); serial RS485 output (optional)Source · 3 ↗ |
| Outputs — Infrared | Analogue 4-20 mA output; 3 relay board; serial RS485 output (optional)Source · 3 ↗ |
| Outputs — PID | Analogue 4-20 mA output; 3 relay board; serial RS485 output (optional)Source · 3 ↗ |
| Maximum power consumption — Catalytic | 5.4 WSource · 2 ↗ |
| Maximum power consumption — Electrochemical | 5.4 WSource · 2 ↗ |
| Maximum power consumption — Intrinsically safe EC | 4.5 WSource · 3 ↗ |
| Maximum power consumption — Infrared | 5.4 WSource · 3 ↗ |
| Maximum power consumption — PID | 5.4 WSource · 3 ↗ |
| Relay contact ratings — Catalytic | 1.0 A at 24 VdcSource · 2 ↗ |
| Relay contact ratings — Electrochemical | 1.0 A at 24 VdcSource · 2 ↗ |
| Relay contact ratings — Infrared | 1.0 A at 24 VdcSource · 3 ↗ |
| Relay contact ratings — PID | 1.0 A at 24 VdcSource · 3 ↗ |
| Display and indication | 7-segment display, 5 status LED and 4 intelligent LEDSource · 2 ↗ |
| Housing material — Catalytic | Enclosure: aluminium pressure die-casting with epoxy paint chromating and chemically resistant Sensor head: stainless steelSource · 3 ↗ |
| Housing material — Electrochemical | Enclosure: aluminium pressure die-casting with epoxy paint chromating and chemically resistant Sensor head: stainless steelSource · 3 ↗ |
| Housing material — Intrinsically safe EC | Enclosure: aluminium with epoxy paint chromating and chemically resistant Sensor head: stainless steelSource · 3 ↗ |
| Housing material — Infrared | Enclosure: aluminium pressure die-casting with epoxy paint chromating and chemically resistant Sensor head: stainless steelSource · 3 ↗ |
| Housing material — PID | Enclosure: aluminium pressure die-casting with epoxy paint chromating and chemically resistant Sensor head: stainless steelSource · 3 ↗ |
| Dimensions — Catalytic | 145 x 187 x 108 mmSource · 3 ↗ |
| Dimensions — Electrochemical | 145 x 187 x 108 mmSource · 3 ↗ |
| Dimensions — Intrinsically safe EC | 138 x 200 x 119 mmSource · 3 ↗ |
| Dimensions — Infrared | 145 x 187 x 108 mmSource · 3 ↗ |
| Dimensions — PID | 145 x 187 x 108 mmSource · 3 ↗ |
| Weight — Catalytic | 1900 gSource · 3 ↗ |
| Weight — Electrochemical | 1900 gSource · 3 ↗ |
| Weight — Intrinsically safe EC | 2700 gSource · 3 ↗ |
| Weight — Infrared | 1900 gSource · 3 ↗ |
| Weight — PID | 1900 gSource · 3 ↗ |
| Storage temperature — Catalytic | -40 / +85°CSource · 3 ↗ |
| Storage temperature — Electrochemical | 0 / +20°C up to 6 monthsSource · 3 ↗ |
| Storage temperature — Intrinsically safe EC | 0 / +20°C up to 6 monthsSource · 3 ↗ |
| Storage temperature — Infrared | -40 / +85°CSource · 3 ↗ |
| Operating temperature — Catalytic | -40 / +60°C *Source · 3 ↗ |
| Operating temperature — Electrochemical | -30 / +55°CSource · 3 ↗ |
| Operating temperature — Intrinsically safe EC | -20 / +50°CSource · 3 ↗ |
| Operating temperature — Infrared | -40 / +60°C *Source · 3 ↗ |
| Relative humidity — Catalytic | 20 / 90% RH @40°C *Source · 3 ↗ |
| Relative humidity — Electrochemical | 15 / 90% RH non condensingSource · 3 ↗ |
| Relative humidity — Intrinsically safe EC | 15 / 90% RH non condensingSource · 3 ↗ |
| Relative humidity — Infrared | 20 / 90% RH @40°C *Source · 3 ↗ |
| Relative humidity — PID | (non condensing)Source · 3 ↗ |
Documented system connections
Related products identified in manufacturer documentation. Check the model and configuration scope in the reference.
- Gas Test Kit · Calibration kitRelationship reference ↗
- STS/CKD+ · Calibration keypadRelationship reference ↗
- STS1REL · Single-relay boardRelationship reference ↗
- STS3REL · Three-relay boardRelationship reference ↗
- ST.S/IDI · RS485 interface boardRelationship reference ↗
- ST.R5M · Remote sensor head kitRelationship reference ↗
- ST.R25M · Remote sensor head kitRelationship reference ↗
Ratings, approvals and compatibility apply only to the models and configurations identified in the manufacturer’s documents.
AL KWARIZMI PROJECT SUPPORTSelection · BOQ review · Technical submittals · Delivery
Technical documents
SMART 3G D2 · Catalytic: Flammable gas (Methane, LPG, Butane, Propane…)↓SMART 3G D2 · Electrochemical cell: Toxic gas (Oxygen, Carbon monoxide, Ammonia…)↓SMART 3G D2 · Electrochemical cell Intrinsically safe: (Chlorine, Sulphur Dioxide, Nitrogen Dioxide)↓SMART 3G D2 · Infrared: Flammable gas, CO2, SF6, Refrigerant gas↓SMART 3G D2 · PID↓SMART 3G D2 · (Post 2021) Flammable gas: Quick guide↓SMART 3G D2 · (Post 2021) Toxic gases intrinsically safe: Manual↓SMART 3G D2 · (Post 2021) Flammable gas: Manual↓SMART 3G D2 · (Pre 2021) Flammable gas: Manual↓SMART 3G D2 · (Post 2021) Toxic gas: Manual↓SMART 3G D2 · (Pre 2021) Toxic gas: Manual↓SMART 3G D2 · (Post 2021) Addendum Display↓SMART 3G D2 · (Pre 2021) Addendum display↓SMART 3G D2 · Kit gas test↓SMART 3G D2 · Keypad STS/CKD+↓SMART 3G D2 · 1 relay STS1REL board↓SMART 3G D2 · 3 relay STS3REL board↓SMART 3G D2 · RS485 board↓SMART 3G D2 · Remote head kit(5m)↓SMART 3G D2 · Remote head kit(25m)↓LET’S BUILD THE RIGHT PACKAGE
From your BOQ.
To a coordinated solution.
Share your scope, drawings and project requirements.